Domain Primer: Cochlear Implants, Audiology, and Digital Audio

Shared technical grounding for PRELUDE. Cochlear implant signal processing is counterintuitive, and plausible-sounding reasoning fails here in ways that are hard to detect by ear. Read this before writing or reviewing DSP code.


1. What a cochlear implant actually does

A cochlear implant (CI) does not amplify sound. It bypasses the entire mechanical and sensory apparatus of the ear and stimulates the auditory nerve directly with electrical pulses. Understanding the signal chain is essential, because a pre-processor can only intervene at one specific point in it.

1.1 The signal chain

acoustic input ──► [ sound processor: external, worn behind ear ] ──► RF coil ──►
   ├─ microphone(s) or Bluetooth/telecoil stream
   ├─ pre-emphasis + front-end AGC / compression
   ├─ analysis filterbank        (12–22 bands)
   ├─ envelope extraction        (rectify + LPF, or Hilbert)
   ├─ channel selection          (n-of-m peak picking, strategy-dependent)
   ├─ loudness mapping           (acoustic dB ──► electrical current units, log compression)
   └─ pulse generation           (biphasic pulse train, interleaved across electrodes)
                                            │
                                            ▼
                     [ implanted receiver/stimulator + electrode array ]
                                            │
                                            ▼
                          spiral ganglion / auditory nerve fibers

The critical architectural fact: everything from the analysis filterbank onward is fixed firmware inside the recipient's processor, tuned by an audiologist. We cannot change it. We can only change the acoustic signal presented to it. This makes the task a pre-compensation (or pre-distortion) problem, exactly analogous to pre-emphasis for a known lossy channel:

Find a transform g such that CI(g(x)) is perceptually closer to the listener's memory of x than CI(x) is.

Note that g is not trying to make g(x) sound good. g(x) will often sound strange or unpleasant to a normal-hearing listener. It is optimized for what survives the CI.

1.2 Tonotopy and the place code

The cochlea is tonotopically organized: high frequencies at the base, low at the apex. The Greenwood function maps characteristic frequency to distance along the cochlea:

F(x) = A · (10^(a·x) − k)        with human values A = 165.4, a = 2.1 (x normalized 0..1), k = 0.88

A typical electrode array is 16–31 mm long and, in practice, is inserted only 18–25 mm into a ~35 mm cochlea. Consequences:

  • The apex is not reached. The most apical electrode typically sits at a place whose natural characteristic frequency is 400–1000 Hz, yet the processor assigns it a band starting near 100–300 Hz. This creates a systematic frequency-place mismatch — the listener hears everything shifted upward in place, which is why CI sound is described as high-pitched, "chipmunk-like", or metallic on initial activation.
  • The brain partially re-maps this over months to years, but never perfectly.
  • Design implication: naive frequency shifting to "fix" the mismatch usually makes things worse in an acclimated user. Their cortex has already adapted to their map. Any frequency warping must be validated perceptually, not assumed.

1.3 The electrical dynamic range — the single most brutal constraint

Acoustic hearing Electrical (CI) hearing
Dynamic range ~120 dB ~6–20 dB (often quoted 6–15 dB)
Just-noticeable intensity steps ~1 dB ~0.5–1 dB, but only ~20 total steps
Frequency channels ~3,500 inner hair cells, ~30 independent critical bands 12–22 electrodes, ~4–8 functionally independent
Temporal fine structure Preserved (phase-locking to ~4–5 kHz) Essentially absent

Each electrode is characterized by two levels set in clinic:

  • THR / T-level: threshold of electrical hearing.
  • MCL / C-level: most comfortable (or maximum comfortable) loudness.

The processor compresses the ~40–60 dB input acoustic range into the T–C range using a logarithmic-ish loudness growth function (often an "instantaneous nonlinear compression" with an adjustable Q-factor / base). Everything the listener hears lives inside that tiny window. This is why heavily compressed, loudness-managed source material tends to survive CI transmission better than wide-dynamic-range material — and why the wide dynamic range of a good classical recording is largely wasted or lost.

1.4 Envelope vs. temporal fine structure (TFS)

Decompose a bandpass signal s(t) via the Hilbert transform:

s(t) = E(t) · cos(φ(t))
       ▲          ▲
    envelope   fine structure (instantaneous phase/frequency)

Standard CI strategies transmit only E(t). The fine structure is discarded and replaced by a fixed-rate electrical pulse carrier. This is the root cause of nearly every CI music deficit:

  • Pitch (F0) is normally coded by TFS phase-locking plus resolved harmonics. With envelope-only coding, pitch must come from (a) coarse place cues and (b) the envelope periodicity rate, which supports pitch only up to ~300 Hz and weakly.
  • Timbre depends on the relative amplitude of resolved harmonics. With 8–12 effective channels, harmonics below ~1.5 kHz are unresolved — several harmonics fall inside one analysis band and are merged into a single envelope value.
  • Melody recognition without rhythm cues is typically near chance for many CI users.
  • Rhythm and tempo are preserved well — envelope modulation is exactly what survives. This is a lever worth exploiting.

1.5 Channel interaction / current spread

Electrical current in the conductive perilymph spreads from the stimulating electrode, decaying roughly exponentially with distance:

I(d) ≈ I₀ · e^(−d/λ)     λ typically corresponds to ~3–8 dB/mm falloff

Adjacent electrodes therefore excite overlapping neural populations. Effects:

  • The effective number of independent channels saturates at ~4–8, regardless of whether the array has 12 or 22 contacts. Adding electrodes past that point yields diminishing returns.
  • Spectral smearing: sharp spectral peaks (formants, harmonics) are blurred.
  • Mitigations in-device: interleaved (non-simultaneous) stimulation — the "S" in CIS — plus current focusing (tripolar/partial-tripolar, phased array).
  • Design implication: deliberately increasing spectral contrast (sharpening peaks, deepening valleys) before transmission can partially pre-compensate for the smearing that follows. This is one of the most promising, evidence-supported levers available to us.

1.6 Coding strategies

Strategy Vendor Mechanism Notes
CIS (Continuous Interleaved Sampling) generic / MED-EL All n channels stimulated every cycle, interleaved Simple, the standard research baseline. What CI_SIM models.
ACE (Advanced Combination Encoder) Cochlear (Nucleus) n-of-m: pick the n highest-envelope of m bands each frame (typ. 8–12 of 22) Default for most Nucleus users. Peak-picking discards low-energy spectral detail.
SPEAK Cochlear n-of-m, lower rate, adaptive n Older.
FSP / FS4 MED-EL Attempts to deliver fine structure on apical channels via zero-crossing-triggered pulse bursts Better F0/music outcomes reported.
HiRes / Fidelity120 / Optima Advanced Bionics High rate + current steering (weighting adjacent pairs to create virtual intermediate channels) Claims more spectral resolution.

Stimulation rate is typically 500–2,400 pulses/s per channel. Higher rate gives finer envelope sampling but more channel interaction and higher power draw.

n-of-m matters enormously for pre-processing. If only the 8 loudest of 22 bands are transmitted per frame, then any spectral content a pre-processor adds that fails to win a peak-picking contest is deleted entirely. Adding energy is not neutral — it can evict other content. Conversely, removing masking content can promote useful content into the transmitted set. This makes n-of-m a competitive, zero-sum allocation problem, and it is a strong argument for source separation and selective de-cluttering rather than additive enhancement.


2. Why music and recorded media are especially hard

CI processors are optimized for real-time speech in moderate noise. Recorded music streamed over Bluetooth violates nearly every assumption:

  1. Polyphony. Multiple simultaneous instruments compete for the same small number of channels. Speech is one source; a mix is many. n-of-m peak-picking under polyphony produces a rapidly fluctuating, incoherent channel selection.
  2. Wide bandwidth. Music has meaningful content from 30 Hz to 16 kHz. The CI's analysis range is roughly 100/200 Hz to 8 kHz. Everything outside is gone.
  3. Wide dynamic range. See §1.3. A 60 dB orchestral crescendo maps into a ~10 dB electrical window.
  4. Pitch and harmony are the content. These are precisely the dimensions the envelope code destroys. Rhythm survives; melody and harmony do not.
  5. Front-end AGC fights the music. The processor's automatic gain control and noise reduction were designed to suppress steady non-speech sounds. Sustained instrumental passages can be treated as noise and attenuated.
  6. Reverb. Reverberant tails fill the envelope valleys between notes and syllables. Because the CI transmits only envelopes, reverb is disproportionately destructive — it directly smears the one cue that survives. Dereverberation is high-value.
  7. Codec interaction. Bluetooth A2DP (SBC/AAC/aptX) is a lossy perceptual codec that discards content it judges masked for a normal-hearing listener. That masking model does not describe a CI listener. Content the codec discards as inaudible may have been exactly the content that would have survived; content it preserves may be irrelevant. Codec artifacts (pre-echo, band truncation) also land in the envelope.

Bluetooth streaming does not bypass the processor's back-end. It bypasses the microphone and (usually) some front-end AGC, but the filterbank, envelope extraction, channel selection, and loudness mapping all still apply. This is what makes pre-processing viable: there is a stable, deterministic transform between the file we control and the percept.


3. Bimodal hearing

A bimodal listener uses a cochlear implant in one ear and an acoustic hearing aid in the other. It is a common configuration, and one with particular relevance to music.

3.1 What the acoustic ear contributes

Residual low-frequency acoustic hearing is disproportionately valuable, because it supplies precisely what the implant cannot:

  • Temporal fine structure below ~1 kHz, giving genuine F0 and pitch perception.
  • Resolved low harmonics, giving timbre, instrument identity, and voice quality.
  • A fundamental frequency to track, which is why melody recognition in bimodal listeners is markedly better than with an implant alone.

The combination is genuinely synergistic — bimodal performance typically exceeds either device alone, and the benefit for music is larger than the benefit for speech.

3.2 What makes bimodal hard

  • Loudness balance between an acoustic and an electric percept must be matched, and drifts as the acoustic ear changes.
  • Frequency-place mismatch across ears: the same acoustic frequency is delivered to different tonotopic places in each ear, impeding binaural fusion.
  • Timing mismatch: hearing-aid processing delay (~5–10 ms) differs from implant processing delay, disrupting interaural time cues.
  • In progressive hearing loss the acoustic ear is a moving target. Every audiogram shift changes the optimal division of labour between the two ears.

3.3 The design constraint this imposes

Where residual acoustic hearing is expected to decline, a system built around it must degrade gracefully to implant-only. Concretely:

  • Do not build anything that structurally depends on residual acoustic hearing.
  • Prefer a two-path architecture — a processing path for the implanted ear and a separate one for the aided ear, with an explicit, adjustable crossover. The system can then be rebalanced as the audiogram changes and eventually collapse to implant-only by setting the acoustic path's contribution to zero. That makes the transition a configuration change rather than a rewrite.
  • Treat any bimodal calibration opportunity as time-limited (see §4).

4. Bimodal listeners as translators between acoustic and electric hearing

Cochlear implant simulations are conventionally validated by playing vocoded audio to normal-hearing listeners. This is a proxy measure, and a weak one: it establishes what a vocoder sounds like to an intact ear, not what an implant sounds like to an implanted one. The direct question is hard to ask, because someone who has only ever heard through an implant has no acoustic reference for comparison, and someone with normal hearing has no implant.

A bimodal listener who acquired hearing loss after developing normal hearing is an unusual and valuable exception. Such a listener

  1. has intact acoustic memory and vocabulary for natural timbre,
  2. currently hears through an implant on one side, and
  3. retains acoustic hearing on the other, permitting genuine within-subject comparison of the same stimulus through two different transduction paths.

This enables a calibration loop unavailable by any other route:

  source audio x
       │
       ├──────────────► aided ear (acoustic reference)  ─┐
       │                                                  ├──► listener reports the difference
       ├──────────────► implanted ear (electric percept) ─┘
       │
       └──► candidate simulation  Ŝ(x) ──► presented to the aided ear only
                                          │
                                          └──► "does this match what the implant sounds like?"
                                                        │
                                                        ▼
                                          iteratively fit simulator parameters

The second loop is the valuable one. Presenting a candidate simulation to the acoustic ear and asking whether it matches the implanted ear's percept of the same source converts an otherwise unfalsifiable model into a fitting problem with a real error signal.

Where hearing loss is progressive, this window closes. Calibration data can only be collected while the listener can still make the comparison, whereas engineering work can be done at any time. That asymmetry should drive scheduling: capture perceptual data early, even before the software is good.

4.1 Elicitation caveats

Subjective auditory report is fragile. Protect the data:

  • Use forced-choice and matching tasks rather than open-ended description wherever possible. "Which of these two is closer?" is testable; "describe what you hear" is useful for generating hypotheses but not for testing them.
  • Randomise and blind presentation order, by script rather than by hand. Include catch trials (identical pairs) to estimate response noise; without a noise floor, a 60% preference is uninterpretable.
  • Beware acclimatisation and learning effects across sessions.
  • Beware anchoring: a listener who has heard a particular simulation many times may begin treating it, rather than the implant, as the reference.
  • Keep sessions short. Auditory fatigue degrades discrimination, and fatigued data is indistinguishable from a null result.
  • Log everything. An unlogged session is lost data, and where the reference ear is declining, the supply of sessions is finite.

5. Levers available to a pre-processor

Ranked roughly by expected value, with the mechanism each exploits:

Lever Mechanism Expected value
Dereverberation Reverb fills envelope valleys; envelopes are all the CI transmits High, well-supported
Source separation / reduce polyphony Fewer competing sources → cleaner n-of-m channel allocation High; demonstrated in CI music literature (simplified/reduced-instrument arrangements are preferred by CI users)
Spectral contrast enhancement Pre-sharpens peaks to counteract current-spread smearing High; direct pre-compensation of a known distortion
Dynamic range compression matched to electrical DR Fits source into the ~10 dB usable window deliberately, rather than letting the processor's generic AGC do it High
Bandwidth remapping / frequency compression Folds musically important content into the 200 Hz–8 kHz transmitted range Medium; risks disrupting an acclimated map — must be validated perceptually
Harmonic / F0 enhancement Strengthens the periodicity cues that weakly survive in the envelope Medium
Transient & onset sharpening Rhythm is the best-preserved dimension; strengthening it plays to the CI's strength Medium, cheap to try
Masker removal (de-cluttering) Under n-of-m, removing a loud irrelevant band promotes a quieter relevant one Medium–high, underexplored
Codec-aware preparation Avoid handing the Bluetooth codec decisions that assume normal-hearing masking Low–medium; hard to control

Anti-lever (do not do): naive loudness maximization / brickwall limiting. It destroys the envelope modulation depth that is the CI's only real information channel.


6. Objective metrics, and why they are not enough

Available instrumental measures:

  • STOI / ESTOI — short-time objective intelligibility; correlates with speech intelligibility, envelope-based, so reasonably relevant to CI.
  • NCM (Normalized Covariance Metric) — envelope-correlation based; among the better predictors for vocoded/CI speech.
  • STMI (Spectro-Temporal Modulation Index) — sensitive to the modulation content that CIs transmit.
  • PESQ / POLQA — designed for telephony codecs; poor fit for CI, use with suspicion.
  • Modulation depth / envelope correlation per channel — a direct, interpretable measure of what actually reaches the nerve. Recommended as the primary internal metric.
  • For music: pitch salience, chroma/melodic contour agreement, spectral centroid and flux distance, onset-detection F1.

None of these are validated against CI percepts, and optimizing them blindly is a known failure mode. They are useful as fast regression signals between listening sessions — to catch outright breakage and to rank candidates before spending scarce listening time. The human loop remains the ground truth. Every metric adopted should be recorded in 05-EVALUATION-PROTOCOL.md along with an honest statement of what it does not capture.


7. Digital audio facts that keep mattering here

  • Nyquist. At a 16 kHz sample rate the maximum representable frequency is 8 kHz. Any filterbank band edge above that is invalid. This is not a hypothetical mistake — see 03-PRIOR-ART.md for a filterbank that silently ran at half its nominal channel count for exactly this reason. A rate of 16 kHz is otherwise reasonable for CI simulation, since the device's own analysis tops out near 8 kHz, but band edges must then be defined relative to 8 kHz rather than to the sample rate. PRELUDE defaults to 20 kHz so that a 300–8500 Hz analysis range is representable.
  • Envelope extraction must be rectify → lowpass (typically 50–400 Hz cutoff) or |hilbert(x)|. Half-wave rectification without a lowpass does not produce an envelope — it produces a distorted copy of the original signal, full of new harmonics.
  • A vocoder needs a carrier. The canonical noise-vocoder is Σᵢ envelopeᵢ(t) · bandpassᵢ(noise) — extract the envelope per band, then multiply it by a band-limited noise (or tone) carrier and sum. Summing rectified band signals is not a vocoder and does not simulate a CI.
  • Analytic signals / Hilbert are the cleanest envelope route, and match what several established reference simulators use.
  • MP3/AAC are lossy and already discard content per a normal-hearing masking model. Prefer lossless sources for research material where possible, so that one perceptual model is not stacked on top of another.
  • Loudness normalization for listening tests must be explicit (target LUFS or matched RMS), otherwise level differences dominate every subjective comparison and invalidate it.

8. Glossary

Term Meaning
ACE Advanced Combination Encoder; Cochlear Ltd's default n-of-m strategy
AGC Automatic Gain Control
C-level / MCL Most comfortable loudness level, per electrode
CI Cochlear Implant
CIS Continuous Interleaved Sampling coding strategy
EAS Electric-Acoustic Stimulation (hybrid: implant + acoustic in the same ear)
F0 Fundamental frequency (perceived pitch of a periodic sound)
Greenwood function Maps cochlear position to characteristic frequency
MAP The individualized program in a CI processor (T/C levels, strategy, rate, allocation)
n-of-m Transmit only the n highest-energy of m analysis bands per frame
pps Pulses per second (stimulation rate)
T-level / THR Threshold level, per electrode
TFS Temporal Fine Structure
Tonotopy Frequency-to-place organization of the cochlea
Vocoder Analysis/resynthesis by band envelopes; the standard CI simulation method

9. References and further reading

Foundational and directly relevant. Entries marked (*) drive design decisions in this codebase and are worth reading before modifying the corresponding stage.

  • Shannon, R. V., Zeng, F.-G., Kamath, V., Wygonski, J., & Ekelid, M. (1995). Speech recognition with primarily temporal cues. Science, 270(5234), 303–304. (*) The founding noise-vocoder paper. Establishes how few channels speech intelligibility requires — and by contrast how many music does.
  • Loizou, P. C. (1998). Mimicking the human ear: an overview of signal-processing techniques for converting sound to electrical signals in cochlear implants. IEEE Signal Processing Magazine, 15(5), 101–130. (*) The standard tutorial on coding strategies.
  • Greenwood, D. D. (1990). A cochlear frequency-position function for several species — 29 years later. JASA, 87(6), 2592–2605. (*) The frequency-to-place map used by the greenwood filterbank spacing.
  • Glasberg, B. R., & Moore, B. C. J. (1990). Derivation of auditory filter shapes from notched-noise data. Hearing Research, 47(1–2), 103–138. (*) ERB scale, used by the erb spacing.
  • Slaney, M. (1993). An efficient implementation of the Patterson–Holdsworth auditory filter bank. Apple Computer Technical Report #35.
  • McDermott, H. J. (2004). Music perception with cochlear implants: a review. Trends in Amplification, 8(2), 49–82. Establishes the rhythm-preserved / pitch-degraded asymmetry that motivates most of §5.
  • Limb, C. J., & Roy, A. T. (2014). Technological, biological, and acoustical constraints to music perception in cochlear implant users. Hearing Research, 308, 13–26.
  • Bierer, J. A. (2010). Probing the electrode–neuron interface with focused cochlear implant stimulation. Trends in Amplification, 14(2), 84–95. Background for the channel interaction model.
  • ITU-R BS.1770-4 (2015). Algorithms to measure audio programme loudness and true-peak audio level. (*) Implemented in prelude.audio.loudness.
  • Bimodal benefit literature — relevant to the two-path crossover design in §3.3; see reviews of combined electric and acoustic stimulation outcomes.

Corrections and additions to this list are welcome; see CONTRIBUTING.md.