The problem

A cochlear implant restores hearing by stimulating the auditory nerve directly. Its sound processor is tuned for real-time speech, and it is very good at that. Recorded music and audiobooks streamed over Bluetooth are a different problem, and the results are widely reported as thin, harsh, or unrecognisable. Music perception in particular remains one of the least satisfying aspects of life with an implant.

The processor itself cannot be changed. Its filterbank, channel selection, loudness mapping, and stimulation parameters are fixed firmware plus a clinical program that only an audiologist should touch. The only available point of intervention is upstream — the audio presented to the device.

That makes this a pre-compensation problem, of the same shape as pre-emphasis for a known lossy channel:

Find a transform g such that CI(g(x)) is perceptually closer to x than CI(x) is.

Note that g(x) need not sound good to a normal-hearing listener. It is optimised for what survives the implant.

What this provides

prelude.ci_sim — a cochlear implant simulator. Models the processor's signal chain stage by stage: analysis filterbank → envelope extraction → n-of-m channel selection → loudness mapping into the electrical dynamic range → channel interaction from current spread → resynthesis. Every intermediate is retained, including the electrodogram — the channel-by-time matrix of stimulation levels, which is what actually reaches the auditory nerve and the right domain in which to compare two signals.

prelude.enhance — pre-processing transforms. (in development)

prelude.audio — I/O and playback safety. Loudness normalisation and true-peak limiting are enforced, not optional. See Safety.

prelude.eval — objective metrics. (in development)

Install

git clone https://github.com/The-Science-Place-Foundation/PRELUDE.git
cd prelude
pip install -e ".[dev,viz]"
pytest

Quick start

from prelude.audio import load_audio, prepare_for_playback, save_audio, Audio
from prelude.ci_sim import SimulatorConfig, simulate

audio = load_audio("input.wav", target_rate=20000)

# An ACE-like device: 8 of 22 channels transmitted per frame.
config = SimulatorConfig(n_channels=22, n_selected=8, seed=0)
result = simulate(audio.samples, audio.sample_rate, config)

print(result.diagnostics())
# {'config_hash': 'e3c1f3d2f972', 'n_channels': 22, 'duration_s': 2.555,
#  'selection_stability': 0.5904, 'mean_channels_active': 8.0,
#  'effective_channels': 16.19}

safe, report = prepare_for_playback(result.audio, audio.sample_rate)
print(report.summary())
save_audio("simulated.wav", Audio(safe, audio.sample_rate),
           metadata={"config_hash": config.hash()})

Or from the command line:

prelude-simulate input.wav -o simulated.wav --channels 22 --selected 8

The simulator reproduces the field's most robust qualitative result — fidelity falls as channel count falls (mean per-band envelope correlation with the source, on a speech sample):

Configuration Envelope correlation with source
CIS, 22 channels 0.83
ACE, 8 of 22 0.75
CIS, 4 channels 0.63

Design principles

Fail loudly on invalid configurations. Band edges above Nyquist raise an error rather than being silently dropped. This is a direct response to a real defect in this project's predecessor, where a nominally 21-channel filterbank ran with 11 channels for months without anyone noticing.

Parameters are data, not code. Configurations live in YAML and carry a hash that is recorded in every output artifact. Parameter sweeps performed by editing source produce results nobody can interpret later.

Every stage is inspectable. CI processing is not intuitively predictable. Being able to look at the electrodogram is how structural mistakes get caught early rather than after months of tuning against a wrong model.

Simulator before enhancer. An enhancer optimised against an inaccurate simulator will confidently produce worse audio. Simulator fidelity gates everything downstream.

Safety

Audio produced by this toolkit may be presented to people whose residual hearing is irreplaceable and, in progressive conditions, already deteriorating. An over-level playback is not a recoverable mistake.

Every signal that may reach a human ear must pass through prelude.audio.loudness.prepare_for_playback, which normalises integrated loudness (default −23 LUFS), enforces a true-peak ceiling (default −1 dBTP), and refuses to process signals containing NaN or infinite values. There is deliberately no bypass argument.

Level matching is also a methodological requirement. Loudness differences dominate every other perceptual judgement, so an unmatched comparison measures level rather than whatever it was intended to measure.

This is research software, not a medical device. It is not certified for clinical use, it must not be used to inform changes to anyone's implant program, and it does not diagnose or treat anything. Device programming is the exclusive province of a qualified audiologist.

Research use

If you are running listening studies with this toolkit, please read docs/05-EVALUATION-PROTOCOL.md first. It covers blinding, catch trials for estimating response noise, loudness matching, session length limits for auditory fatigue, and the reasons each matters.

Human-subjects research requires ethics review. Studies involving a small number of participants — including single-case designs, which are legitimate and well-established methodology — still require it, and any conflict of interest between investigator and participant must be disclosed.

Documentation

Document Purpose
Domain primer CI physiology, audiology, and DSP grounding — read before writing DSP
Architecture System design and stage contracts
Evaluation protocol Metrics and listening-study protocol
Prior art Predecessor work, and the defects this design avoids
Decisions Architecture Decision Records

Project status

Alpha. The simulator core is implemented and tested. The enhancement and evaluation modules are scaffolded but incomplete. APIs will change.

The simulator currently models a generic implant. Manufacturer-specific frequency allocation tables, T/C levels, and front-end processing are supported by the configuration format but are not bundled — supply your own via config/device_profile.yaml.

The simulator has not yet been validated against an external reference. It is internally consistent, passes its behavioural test suite, and reproduces the channel-count/fidelity relationship that defines CI simulation — all necessary conditions, none sufficient. A first comparison against an independent implementation showed substantial disagreement that is not yet explained; the investigation is written up in docs/lab-notebook/2026-07-25-reference-simulator-comparison.md. Do not treat simulator output as an accurate model of any real device.

Contributing

Contributions are very welcome, particularly from audiologists, CI users, and hearing researchers. Domain corrections are as valuable as code. See CONTRIBUTING.md.

About

PRELUDE is developed by The Science Place Foundation, a 501(c)(3) nonprofit, and released as open source for the audiology and hearing-research community.

The project began as an attempt to solve a specific problem for one bimodal cochlear implant user: recorded music and audiobooks, streamed to an implant, sound markedly worse than the same material heard acoustically. That problem is not specific to one person, and neither is the approach. The toolkit is built to be useful to any researcher or clinician working on cochlear implant audio.

If you are working in this area and PRELUDE is missing something you need, please open an issue — see CONTRIBUTING.md.

Acknowledgements

This work builds on the noise-vocoder simulation method introduced by Shannon et al. (1995), Speech recognition with primarily temporal cues, Science 270(5234).

PRELUDE is an independent implementation of published methods. Every stage is built from the primary literature cited in docs/01-DOMAIN-PRIMER.md; no third-party simulator code is used, derived from, or redistributed here.

Two existing simulators served as black-box references during development, and their published documentation clarified parameter semantics: the Cochlear Implant Simulation package from the Universidad de Granada (de la Torre Vega, Bastarrica Martí, de la Torre Vega & Sainz Quevedo, 2004), and a MATLAB CI demonstration by Amir Rahimzadeh (HTW Berlin, 2013) built on Malcolm Slaney's Auditory Toolbox. We are grateful to both groups for publishing their tools and documentation.

Citation

If PRELUDE contributes to published work, please cite it — see CITATION.cff, which GitHub renders as a "Cite this repository" link.

Licence

GNU Affero General Public License v3.0 — see LICENSE.

This is a strong copyleft licence, chosen deliberately. In plain terms:

  • You must give credit. Copyright and licence notices have to be retained.
  • You may use this for anything, including commercially and in clinical settings — there is no field-of-use restriction, because excluding manufacturer-funded and for-profit clinical research would exclude much of the field this work is meant to serve.
  • Derivatives must stay open. If you modify PRELUDE and distribute it, your modifications must be released under the AGPL as well.
  • Network use counts as distribution. If you run a modified version as a hosted service, you must offer its source to the users of that service. This is what distinguishes the AGPL from the ordinary GPL, and it closes the loophole whereby a company builds on the work and never ships a binary.

The intent is that nobody can take this work, build a proprietary product on top of it, and give nothing back — while keeping the toolkit genuinely open for researchers, clinicians, and implant users.

Apache-2.0 has a patent grant that the AGPL lacks in equivalent form, which is a real consideration in a patent-dense field. The AGPL's §11 does include a patent grant from contributors, but it is narrower. Contributions are accepted on the basis that contributors licence their own patent claims accordingly.

Copyright © The Science Place Foundation and the PRELUDE contributors.