The software tools used by service personnel to test &...

Critical severity Unreviewed Published Nov 15, 2024 to the GitHub Advisory Database • Updated Nov 15, 2024

Package

No package listedSuggest a package

Affected versions

Unknown

Patched versions

Unknown

Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

References

Published by the National Vulnerability Database Nov 14, 2024
Published to the GitHub Advisory Database Nov 15, 2024
Last updated Nov 15, 2024

Severity

Critical

CVSS overall score

This score calculates overall vulnerability severity from 0 to 10 and is based on the Common Vulnerability Scoring System (CVSS).
/ 10

CVSS v3 base metrics

Attack vector
Network
Attack complexity
Low
Privileges required
None
User interaction
None
Scope
Changed
Confidentiality
High
Integrity
High
Availability
High

CVSS v3 base metrics

Attack vector: More severe the more the remote (logically and physically) an attacker can be in order to exploit the vulnerability.
Attack complexity: More severe for the least complex attacks.
Privileges required: More severe if no privileges are required.
User interaction: More severe when no user interaction is required.
Scope: More severe when a scope change occurs, e.g. one vulnerable component impacts resources in components beyond its security scope.
Confidentiality: More severe when loss of data confidentiality is highest, measuring the level of data access available to an unauthorized user.
Integrity: More severe when loss of data integrity is the highest, measuring the consequence of data modification possible by an unauthorized user.
Availability: More severe when the loss of impacted component availability is highest.
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Weaknesses

CVE ID

CVE-2024-48966

GHSA ID

GHSA-285j-vfp9-5w9w

Source code

No known source code

Dependabot alerts are not supported on this advisory because it does not have a package from a supported ecosystem with an affected and fixed version.

Learn more about GitHub language support

Loading Checking history
See something to contribute? Suggest improvements for this vulnerability.